Doctor/Nurse Calling System
Application :
In any Hospital it is essential to facilitate Patients to
call the Duty Doctor/Staff Nurse/House-Keepers for
attention /assistance during emergencies.
It is desirous that the ‘Doctor/Nurse Calling System’
be ‘Fool-Proof’ & attention to Patients is 100%

Technical Specifications
Operate Volts
Choice of
Windows per
DCU

85 to 275 V AC or DC
Min :
Max :

08 Windows per DCU
Any No. of Windows per DCU

Individual
Window Size

70mm x 35mm

‘Doctor/Nurse Calling’ Unit [DCU] is usually mounted

Operate Temp

0° C to 70° C, Upto 95% RH

in a ‘Centralised-Area’, say ~ the ‘Duty-Room’ of

Storage Temp

-10° C to 90° C, Upto 95% RH

Burden

2.5 Watts @ 230 VAC [Max]
per DCU Window

Choice of
Sequences of
Operation

[1]. Auto-RESET [AR-01]
[2]. Manual-RESET [MR-01 & MR-02]
[3]. Ringback [RBM-01]

Choice of
Switch Types

[1]. Regular ‘ON/OFF’ [Toggle type]
[2]. ‘Calling-Bell’ [Spring-Return type]

assured, beyond ambiguity.

Doctors/Nurses, in each floor of a Hospital.


DCU announces the ‘Call from Patient’ by means of
‘Visual’ & ‘Audible’ Alarms, to alert the Duty
Doctors/Nurses and to initiate remedial action.



For ‘Visual’ Alarm, ‘Rear-Illuminated Windows’ are
provided. Patient’s Room/Bed No. is inscribed on
Front-side of Window.



For ‘Audible’ Alarm, in-built ‘Buzzer’ is provided.

Description :

… [Default size]

Refer : Sequences of Operation [overleaf]
for details of ‘Recommended Switch types’

‘Doctor/Nurse Calling’ Unit [DCU] consists of :

DEDEDE

1. “Windows” to indicate Patient’s Room / Bed No.
One Window for each Room/Bed is provided.
2. Push Buttons for ‘ACCEPT’, ‘Lamp TEST’ & MUTE
3. Built-in Audible Alarm.
4. Each Room/Bed is provided with one ‘Call-Switch’
for use by the Patient
M e t h o d o f O p e r a t i o n : {Auto-RESET [AR-01]}


To call Doctor/Nurse, Patient has to Press the
Switch provided by his/her Bed-side.



‘Rear-Illuminated Window’ displaying Patient’s
Room/Bed No., starts ‘Flashing’ & Audible Alarm

Ordering Information
Following ‘Ordering Information’ to be specified by
User, beyond ambiguity, for Enquiry or Purchase Order
No. of Windows
per DCU
required
Sequence of
Operation
required

_____ No. of Windows / DCU
______________________
[specify any ONE from above THREE]

switches ON, thus alerting the Doctor/Nurse.
Doctor/Nurse should press ‘ACCEPT’ Push Button

DEDEDE

to acknowledge ‘Call’ from the Patient.
‘Flashing’ Window become ‘Steady’ glow & Audible
Alarm switches OFF.
Doctor/Nurse should go to Patient’s Room/Bed,
attend to his/her needs and before returning
should switch ‘OFF’ the Patient’s bed-side ‘CallSwitch’ to ‘RESET’ the ‘Call’ from Patient.
As soon as ‘Call-Switch’ is switched OFF, ‘Steady’
glow of Window on DCU is switched OFF, thereby

for further details, please contact :

Bharani Electronics
Door # 27, 46th Street, Ashok Nagar,
Chennai : 600 083. India.
Phone/Fax : [044] 2489 5421 / 2489 1226
E-Mail : [email protected]

completing the episode.
Product improvement is a continuous feature.
Specifications / Design listed in this Publication are therefore subject to revision, without notice.

Choice of Sequences of Operation
Sequence Designation

Auto-RESET

AR-01

Line
No

Initiating
Contact

Push
Button
Pressed

1

Normal

--

OFF

OFF

--

2

Abnormal

--

Flash

ON

Lock-in

3

Abnormal

ACCEPT

Steady

OFF

-Auto-RESET
acted

Visual
Alarm

Audible
Alarm

Remarks

4

Normal

--

OFF

OFF

5

Normal before
ACCEPT

--

Flash

ON

Lock-in

OFF

Auto-RESET
acted

6

Normal

RESET

Regular ‘ON/OFF’

Recommended Switch type

Manual-RESET

OFF

[Toggle type]

ONLY

Sequence Designation

Sequence Designation

MR-01

MR-02

Line
No

Initiating
Contact

Push
Button
Pressed

1

Normal

--

OFF

OFF

--

OFF

OFF

--

2

Abnormal

--

Flash

ON

Lock-in

Flash

ON

Lock-in

3

Abnormal [or]
Normal

ACCEPT

Steady

OFF

Steady

OFF

Manual RESET
required

4A

Abnormal

RESET

Flash

ON

Manual RESET
required
To Line 2,
ACCEPT reqd
Manual RESET
acted

Steady

OFF

To Line 3

OFF

Manual RESET
acted

4B

Normal

RESET

Visual
Alarm

Audible
Alarm

Remarks

Visual
Alarm

Audible
Alarm

Remarks

OFF

Recommended Switch type

OFF

OFF

[1]. Regular ‘ON/OFF’ [Toggle type] … or …
[2]. ‘Calling-Bell’ [Spring-Return type]

Sequence Designation

Ringback

RBM-01

Line
No

Initiating
Contact

Push Button
Pressed

Visual
Alarm

Audible
Alarm

Ringback
Audible Device

1

Normal

--

OFF

OFF

OFF

2

Abnormal

--

Fast Flash

ON

OFF

3

Abnormal

ACCEPT

Steady

OFF

OFF

4

Normal

--

Slow Flash

OFF

ON

5

Abnormal

--

<<<<<

To Line 2

>>>>>

6

Normal

RESET

OFF

OFF

OFF

7

Normal before ACCEPT

--

Fast Flash

ON

OFF

8

Normal

ACCEPT

Slow Flash

OFF

ON

9

Abnormal

--

<<<<<

To Line 2

>>>>>

10

Normal

RESET

OFF

OFF

OFF

11

Normal

TEST

Slow Flash

OFF

ON

Recommended Switch type

[1]. Regular ‘ON/OFF’ [Toggle type] … or …
[2]. ‘Calling-Bell’ [Spring-Return type]