Doctor/Nurse Calling System
Application :
In any Hospital it is essential to facilitate Patients to
call the Duty Doctor/Staff Nurse/House-Keepers for
attention /assistance during emergencies.
It is desirous that the ‘Doctor/Nurse Calling System’
be ‘Fool-Proof’ & attention to Patients is 100%
Technical Specifications
Operate Volts
Choice of
Windows per
DCU
85 to 275 V AC or DC
Min :
Max :
08 Windows per DCU
Any No. of Windows per DCU
Individual
Window Size
70mm x 35mm
‘Doctor/Nurse Calling’ Unit [DCU] is usually mounted
Operate Temp
0° C to 70° C, Upto 95% RH
in a ‘Centralised-Area’, say ~ the ‘Duty-Room’ of
Storage Temp
-10° C to 90° C, Upto 95% RH
Burden
2.5 Watts @ 230 VAC [Max]
per DCU Window
Choice of
Sequences of
Operation
[1]. Auto-RESET [AR-01]
[2]. Manual-RESET [MR-01 & MR-02]
[3]. Ringback [RBM-01]
Choice of
Switch Types
[1]. Regular ‘ON/OFF’ [Toggle type]
[2]. ‘Calling-Bell’ [Spring-Return type]
assured, beyond ambiguity.
Doctors/Nurses, in each floor of a Hospital.
•
DCU announces the ‘Call from Patient’ by means of
‘Visual’ & ‘Audible’ Alarms, to alert the Duty
Doctors/Nurses and to initiate remedial action.
•
For ‘Visual’ Alarm, ‘Rear-Illuminated Windows’ are
provided. Patient’s Room/Bed No. is inscribed on
Front-side of Window.
•
For ‘Audible’ Alarm, in-built ‘Buzzer’ is provided.
Description :
… [Default size]
Refer : Sequences of Operation [overleaf]
for details of ‘Recommended Switch types’
‘Doctor/Nurse Calling’ Unit [DCU] consists of :
DEDEDE
1. “Windows” to indicate Patient’s Room / Bed No.
One Window for each Room/Bed is provided.
2. Push Buttons for ‘ACCEPT’, ‘Lamp TEST’ & MUTE
3. Built-in Audible Alarm.
4. Each Room/Bed is provided with one ‘Call-Switch’
for use by the Patient
M e t h o d o f O p e r a t i o n : {Auto-RESET [AR-01]}
•
To call Doctor/Nurse, Patient has to Press the
Switch provided by his/her Bed-side.
•
‘Rear-Illuminated Window’ displaying Patient’s
Room/Bed No., starts ‘Flashing’ & Audible Alarm
Ordering Information
Following ‘Ordering Information’ to be specified by
User, beyond ambiguity, for Enquiry or Purchase Order
No. of Windows
per DCU
required
Sequence of
Operation
required
_____ No. of Windows / DCU
______________________
[specify any ONE from above THREE]
switches ON, thus alerting the Doctor/Nurse.
Doctor/Nurse should press ‘ACCEPT’ Push Button
DEDEDE
to acknowledge ‘Call’ from the Patient.
‘Flashing’ Window become ‘Steady’ glow & Audible
Alarm switches OFF.
Doctor/Nurse should go to Patient’s Room/Bed,
attend to his/her needs and before returning
should switch ‘OFF’ the Patient’s bed-side ‘CallSwitch’ to ‘RESET’ the ‘Call’ from Patient.
As soon as ‘Call-Switch’ is switched OFF, ‘Steady’
glow of Window on DCU is switched OFF, thereby
for further details, please contact :
Bharani Electronics
Door # 27, 46th Street, Ashok Nagar,
Chennai : 600 083. India.
Phone/Fax : [044] 2489 5421 / 2489 1226
E-Mail :
[email protected]
completing the episode.
Product improvement is a continuous feature.
Specifications / Design listed in this Publication are therefore subject to revision, without notice.
Choice of Sequences of Operation
Sequence Designation
Auto-RESET
AR-01
Line
No
Initiating
Contact
Push
Button
Pressed
1
Normal
--
OFF
OFF
--
2
Abnormal
--
Flash
ON
Lock-in
3
Abnormal
ACCEPT
Steady
OFF
-Auto-RESET
acted
Visual
Alarm
Audible
Alarm
Remarks
4
Normal
--
OFF
OFF
5
Normal before
ACCEPT
--
Flash
ON
Lock-in
OFF
Auto-RESET
acted
6
Normal
RESET
Regular ‘ON/OFF’
Recommended Switch type
Manual-RESET
OFF
[Toggle type]
ONLY
Sequence Designation
Sequence Designation
MR-01
MR-02
Line
No
Initiating
Contact
Push
Button
Pressed
1
Normal
--
OFF
OFF
--
OFF
OFF
--
2
Abnormal
--
Flash
ON
Lock-in
Flash
ON
Lock-in
3
Abnormal [or]
Normal
ACCEPT
Steady
OFF
Steady
OFF
Manual RESET
required
4A
Abnormal
RESET
Flash
ON
Manual RESET
required
To Line 2,
ACCEPT reqd
Manual RESET
acted
Steady
OFF
To Line 3
OFF
Manual RESET
acted
4B
Normal
RESET
Visual
Alarm
Audible
Alarm
Remarks
Visual
Alarm
Audible
Alarm
Remarks
OFF
Recommended Switch type
OFF
OFF
[1]. Regular ‘ON/OFF’ [Toggle type] … or …
[2]. ‘Calling-Bell’ [Spring-Return type]
Sequence Designation
Ringback
RBM-01
Line
No
Initiating
Contact
Push Button
Pressed
Visual
Alarm
Audible
Alarm
Ringback
Audible Device
1
Normal
--
OFF
OFF
OFF
2
Abnormal
--
Fast Flash
ON
OFF
3
Abnormal
ACCEPT
Steady
OFF
OFF
4
Normal
--
Slow Flash
OFF
ON
5
Abnormal
--
<<<<<
To Line 2
>>>>>
6
Normal
RESET
OFF
OFF
OFF
7
Normal before ACCEPT
--
Fast Flash
ON
OFF
8
Normal
ACCEPT
Slow Flash
OFF
ON
9
Abnormal
--
<<<<<
To Line 2
>>>>>
10
Normal
RESET
OFF
OFF
OFF
11
Normal
TEST
Slow Flash
OFF
ON
Recommended Switch type
[1]. Regular ‘ON/OFF’ [Toggle type] … or …
[2]. ‘Calling-Bell’ [Spring-Return type]